Herunterladen Inhalt Inhalt Diese Seite drucken

Configuring The Measurement - +Gf+ 2298 Bedienungsanleitung

Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Type 2298 80 GHz Radar Level Sensor
5.2

Configuring the measurement

5.2.1
P00 Unit system, default units, regional parameter
FACTORY DEFAULT: 0000
If parameter P00 is changed, the device reverts the entire parameter set to the factory default values of the new unit system.
Therefore, all parameters must be set again!
P00
a
Application ➙ Operating mode
EView2
a
Mode
0
Normal
1
High-sensitivity
P00
b
Application ➙ Engineering units
EView2
b
Unit
Metric (EU)
0
m
1
cm
2
mm
P00
c
Application ➙ Calculation system
EView2
c
Region / Unit system
0
EU / Metric
1
US / Imperial
2
Region 2 / Metric
3
Region 2 / Imperial
4*
Region 3 / Metric
5*
Region 4 / Metric
P00
d
Measurement configuration ➙ Temperature
EView2
d
Region / Unit system
0
°C
1
°F
24
Imperial (US)
ft
inch
inch
Regional parameter
EU, United Kingdom,
Albania, Andorra, Azerbaijan,
Australia, Belarus, Bosnia and
Herzegovina, Canada,
Liechtenstein, Moldavia,
Monaco, Montenegro, New
Zealand, Northern Macedonia,
Norway, San Marino, Saudi
Arabia, Serbia, Switzerland,
Turkey, Ukraine
USA
Brazil, Japan, South Korea,
Taiwan, Thailand
India, Malaysia, South Africa
Russia, Kazakhstan
* The accuracy of ±2 mm is not guaranteed for
Region 3 and Region 4 settings.