Naída CI Q90 is properly connected to the AIM Programming Cable
o
Close and relaunch the OM Suite application.
o
Close the OM Suite application, restart the AIM System, and relaunch the OM Suite
application.
•
Problems with acoustic output. Confirm or perform the following:
o
Insert Earphone is properly connected to the cable.
o
Insert Earphone cable is connected to the AIM System.
o
AIM Sterile Insert tube is properly connected to the AIM Insert Earphone.
o
Perform system verification as described in the Maintenance section of these instruc-
tions.
o
Close and relaunch the OM Suite application.
o
Close the OM Suite application, restart the AIM System, and relaunch the OM Suite
application.
•
If the OM Suite application crashes restart the AIM System and relaunch the application.
Guidance & Manufacturer's Declarations
Per IEC 60601-1-2
Electromagnetic emissions
Emissions Test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
Voltage fluctuations/ flicker emissions
Immunity
The AIM System is intended for use in the electromagnetic environment
specified below. The user of the AIM System should assure that it is used in such
an environment.
Immunity
ESD
Power frequency magnetic field (50 & 60 Hz)
Radiated RF
Electrical fast transient/burst (100 kHz repe-
tition)
IEC 61000-4-4
Surges line-to-ground
IEC 61000-4-5
24
Compliance
Group 1
Class A
IEC 61000-3-2
IEC 61000-3-3
Test level
±8 kV contact; ±15 kV air ±8 kV contact; ±15 kV air
30 A/m
3 V/m
80 MHz to 2.7 GHz
±1 kV
100 kHz repetition rate
±2 kV
Compliance level
30 A/m
3 V/m
±1 kV
±2 kV