3.5. Replacement of lamp

ri mini®
Remove instrument head from battery handle.
The lamp is in the bottom section of the instrument head.
Remove lamp, by using your thumb and forefinger or a suitable
tool, from instrument head. Firmly insert new lamp.
pen scope®
Screw off the metal fitting. Release lamp by turning in
counter-clockwise direction.
Insert new lamp by screwing in in clockwise direction
and screw on the metal fitting again.
3.6. Specification of lamp
ri-mini
otoscope
®
pen-scope
®
3.7. Care
The otoscope may be cleaned externally with a moist cloth.
Cleaning and/or disinfection
In addition, the following disinfectants may be used for disinfec-
tion:
Aldehyde (formaldehyde, glutaraldeyhde, aldehyde derivatives),
or surfactants.
In addition, all instrument components, with the exception of the
magnifying glass, may be disinfected with alcohol.
When using these substances, the manufacturer's instructions
must be strictly complied with.
Means for cleaning or disinfection may be a soft, possibly lint-
free cloth or Q-tips.
Never immerse the otoscope in fluid.
Warning
Ensure that no fluid penetrates into the interior of the enclosure.
Sterilisation
According to the ruling teaching (Tübingen Medical Products
a) Otoscope
Test
Centre), sterilisation is only specified for surgery. Due to the
otoscope not being used in surgery, cleaning or disinfecting will
be fully adequate.
Although sterilisation is not required, as specified in a), the ear
b) Reusable ear specula
speculum may nevertheless be sterilised at 134° C for a dwell
time of 10 minutes in a steam steriliser.
otoscope
2,5 V
average life 15 h
2,7 V/HL2,5 V
average life 10 h
750 mA
300 mA
13