Table 5: Manufacturer's Declaration
Proximity Fields
The Vercise Genus DBS System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The users
of the Vercise Genus DBS System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Vercise Genus DBS System as recommended below, according to the maximum output power
of the communications equipment.
Proximity Test
IEC 60601‑1‑2 Test Level
385 MHz: 27 V/m @ 18 Hz pulse modulation
450 MHz: 28 V/m @ FM modulation
IEC 61000-4-3
710 MHz, 745 MHz, 780 MHz: 9 V/m @ 217 Hz pulse modulation
1720 MHz, 1845 MHz, 1970 MHz: 28 V/m @ 217 Hz pulse modulation
1720 MHz, 1845 MHz, 1970 MHz: 28 V/m @ 217 Hz pulse modulation
2450 MHz: 28 V/m @ 217 Hz pulse modulation
IEC 61000-4-3
5240 MHz, 5500 MHz, 5785 MHz: 9 V/m @ 217 Hz pulse modulation
Note: For the frequency bands in this table, use the recommended separation distance. The recommended minimum separation distance of 30 cm
between portable and mobile RF communications equipment (transmitters) and the Vercise Genus DBS System apply to all other frequencies within
the stated ranges.
Note: These guidelines may not apply in all situations. Absorption and reflection from structures, objects, and people affects electromagnetic
propagation.
Information for Patients
92694767-14
36 of 255
Electromagnetic
Compliance Level
Environment Guide
27 V/m
Recommended
28 V/m
separation distance
9 V/m
d = 30 cm
28 V/m
28 V/m
Recommended
28 V/m
separation distance
d = 30 cm
9 V/m