IMPAC pyrometers ISR 12-LO · IGAR 12-LO

Legend

Note: The note symbol indicates tips and useful information in this manual.
All notes should be read with regard to an effective operation of the instrument.
Security note laser beam
Indicates to the danger of a built-in laser targeting light.
MB
Shortcut for Temperature range (in German: Messbereich)

Terminology

The used terminology corresponds to the VDI- / VDE-directives 3511, page 4.

Disposal / decommissioning

Inoperable IMPAC pyrometers have to be disposed corresponding to the local regulations of electro or elec-
tronic material.
1

Technical data

Temperature ranges:
Sub range:
Signal processing:
Spectral ranges:
Measuring modes:
Power supply:
Power consumption:
Analog output:
Digital interface:
Resolution:
Isolation:
Display:
Parameters:
Emissivity slope K:
ε
Emissivity
:
Switch-off limit:
Exposure time t
:
90
4
ISR 12-LO:
MB 13:
MB 18:
MB 25:
MB 33:
IGAR 12-LO: MB 10:
MB 13:
MB 22:
any range adjustable within the temperature range, minimum span 51°C
photoelectric current, digitized
λ
ISR 12-LO:
: 0.8 µm /
1
λ
IGAR 12-LO:
: 1.52 µm /
1
Ratio- (2-color-) / mono- / metal mode switchable
24 V DC (15 to 40 V DC) or 24 V AC (12 to 30 V AC), 48 to 62Hz
max. 10 W
0 to 20 mA or 4 to 20 mA switchable, load 0 to 500 Ohm
Test current 10 mA by pressing test key
Switchable: RS232 or RS485 addressable (half duplex),
baud rate 2.4 up to 115.2 kBd
Display: 1°C
Interface: 0.1°C
Analog output: < 0.025% of temperature range
power supply, analog output and digital interface are galvanically isolated
from each other
Built-in 4-digit 7-segment-LED, height 13 mm; LED's for:
°C/°F; clear mode "auto","ext"; ratio- (2-color) / mono- / metal mode
Adjustable or readable at the device or via interface:
ε
Emissivity
, emissivity slope K, exposure time t
mum value storage t
CL
value storage, switching measuring mode, switch-off limit, warning level
dirty window, 0 to 20 or 4 to 20 mA, sub range, address, baud rate, °C/°F
Readable at the device or via interface:
Measuring temperature
Adjustable only via interface commands:
wait time t
, keyboard lock
W
ε
ε
/
: 0.8 to 1.2 adjustable in steps of 0.001
1
2
0.1 to 1 adjustable in steps of 0.001
2% to 50% in 1%-steps
2 ms (with dynamical adaptation at low signal levels),
adjustable at 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
600 to 1300°C
750 to 1800°C
900 to 2500°C
1000 to 3300°C
300 to 1000°C
350 to 1300°C
500 to 2200°C
λ
: 1.05 µm
2
λ
: 1.64 µm (MB 22:
2
, automatically or external deletion of maximum
λ
λ
: 1.28 /
: 1.65 µm)
1
2
, clear times for maxi-
90