14. Autoclaving

1. The entire pipette adapter
incl. membrane filter is
autoclavable at 121°C
(249,8 °F) (2 bar) according
to DIN EN 285.
Note:
Not every membrane filter is suit-
able. Only use genuine manufac-turer's
recommended accessories.
Each instrument will be supplied
with a 0.2 µm membrane filter (red
color code). The membrane filter is
autoclavable up to 5 times at 121 ºC
(249,8 °F). It is the user's responsibility
to ensure effective autoclaving. The filter
with 0.2 µm pore size (for cell culture)
is sufficient to meet class B tolerances.
For class A accuracy, it is necessary to
use a 3 µm membrane filter. This filter
has a larger pore size and thus permits
better airflow for gravity-delivery (Spare
parts page 24).
2. Reassemble the instrument in reverse
order and carry out leak test.

14. Autoklavieren

1. Die komplette
Pipettenaufnahme
inkl. Membranfilter ist
autoklavierbar bei 121 °C
(249,8 °F) (2 bar) nach DIN
EN 285.
Hinweis:
Nicht jeder Membranfilter ist geeig-
net! Nur Original-Zubehör verwen-den.
Serienmäßig ist das Gerät mit
0,2 µm Membranfilter (roter Farb-
code) ausgestattet. Der Membran-filter
ist bis zu fünfmal bei 121 ºC
(249,8 °F) autoklavierbar. Die
Wirksamkeit des Autoklavierens ist vom
Anwender jeweils selbst zu prüfen. Mit
0,2 µm Membranfilter (für Zellkultur)
können Volumentoleranzen der Klasse
B erzielt werden. Zur Ein-haltung
der Volumentoleranzen der Klasse A
sind 3 µm Membranfilter erforderlich.
Diese haben aufgrund der größeren
Porenweite einen geringeren Einfluss
auf die freie Ablaufzeit der Pipetten
(Ersatzteile S. 24).
2. Gerät in umgekehrter Reihenfolge
wieder komplettieren und Dichtigkeit
prüfen.