11. RUN a HbA1c TEST
STEP 6.
Insert the reagent pack in the cartridge and gently push into cartridge compartment
of the analyzer. The icon 'Close the lid' will be displayed.
NOTE: Do not force the reagent pack into the cartridge, it will only fit one
way.
STEP 7.
The test starts automatically when the lid is closed.
IMPORTANT: Do not open the lid during testing.
Do not move the analyzer during testing.
Do not expose the analyzer to vibration during testing.
STEP 8.
The HbA1c result displays after 5 minutes.
14
11. RUN a HbA1c TEST
NOTE: If the results ">14%" or "<4%" are displayed repeat testing to confirm
result. If the second result also is outside the range, contact your local
distributor.
STEP 9.
After the test is completed, open the analyzer lid. The display will show 'Remove
Cartridge' . Take out the cartridge from the analyzer by gently pushing it to the left
and pulling it out.
CAUTION: Do not force the cartridge to remove it from the analyzer.
Dispose all waste in accordance with applicable national and/or local
regulations.
Code confirmation display after testing
If an error in the cartridge code recognition process occurs, the analyzer will ask for
type and code of the cartridge.
1.Take out the cartridge, use
and
to match the type of cartridge, press
to confirm.
2. After selecting the type of cartridge, use
and
to match the code
number of the cartridge, press
confirm, and the test result is shown.
15
to