17. SPECIFICATION

Sample Type
Capillary whole blood,
Venous whole blood with anticoagulant
Sample Volume
4 μL
Test Range
4.0-14.0%(NGSP), 20-130mmol/mol(IFCC)
Reading Time
5 minutes
Memory Capacity
200 test results
Wavelength
415 nm
Power Required
DC 9 V-1.5 A
Dimensions/Weight
198*217*136(mm) / 1.6kg
Storage Temperature Analyzer
10 ~ 35°C (50 ~ 95°F)
Operating Temperature
17 ~ 32°C (63 ~ 90°F)
Relative Humidity Range
10% ~ 90%
Options
Print support – conncection to Thermal Printer
for printing support.
Barcode scanner for scanning of patient ID
operator ID.
PC application for transfer of data to computer.

18. MAINTENANCE

Recommended maintenance routine
1) Exterior cleaning of the analyzer
2) Cartridge compartment cleaning
3) Replacing air filter
34
18. MAINTENANCE
Cleaning the analyzer
1) Disconnect the analyzer before cleaning.
2) Wipe off the Analyzer using a soft, clean cloth.
3) Do not use liquid or aerosol cleaners.
4) If necessary, an alcohol swab can be used to remove dirt, other chemical
solutions should not be used. Cleaning solutions may cause damage to the
analyzer surface.
Cleaning the cartridge compartment of the analyzer
STEP 1. Turn the power off and disconnect the power cord before cleaning.
STEP 2. When the lid is opened,
press the two black knobs to rotate
the cartridge compartment forward.
STEP 3. Moisten a sponge with ethanol or recommended commercial disinfectant
(Recommended disinfectants: CaviCide
STEP 4. Gently wipe off the cartridge compartment using the sponge.
Do not to drip any liquids into the system.
Replacing the Fan Filter:
Regularly check the fan filters located on the back of the analyzer. Replace
the fan filter if clogged by dust. Extra filters are included at delivery of the analyzer.
STEP 1. Remove
the fan filter cover.
Filter cover
NOTE: Additional fan filters can be purchased from your local distributor.
®
®
, METREX
RESEARCH COPORATION, US)
STEP 2. Take out
the used fan filter -
replace with a new.
STEP 3. Close the
cover.
Filter
35