11. RUN a HbA1c TEST
Expected values
ADA's most recent Clinical Practice Recommendation for diabetes specifies a
treatment goal of less than 7.0 HbA1c %, 53 mmol/mol.
[Reference]
The American Diabetes Association's (ADA's) 2012 Clinical Practice.
Recommendation for diabetes specifies a treatment goal of less than 7 HbA1c %, 53
mmol/mol.
Limitation of Procedure
®
The HemoCue
HbA1c 501 assay gives accurate and precise hemoglobin results in
the range 7 to 20 g/dL. Most patients will have hemoglobin concentrations within
this range.
However, patients with severe anemia may have hemoglobin concentrations below
7 g/dL, and patients with polycythemia may have hemoglobin concentrations above
20 g/dL. Patients known to have these conditions should be tested with another
method for determination of HbA1c%.
12. REVIEWING RESULTS
STEP 1.
In stand-by mode, press
button for 3 seconds for
Press
or
button to select Data, then press
STEP 2.
The test results will appear chronologically starting with the most recent date.
Press
and
buttons to scroll through the test results.
or
.
.
16
13. SYSTEM CHECK
Quality Control
®
The HemoCue
HbA1c 501 Check Cartridge screens the optical and operating
systems of the Analyzer.
Type of Check Cartridges
®
•
HemoCue
HbA1c 501 Daily Check Cartridge
®
•
HemoCue
HbA1c 501 Monthly Check Cartridge
Storage Instruction
The Check Cartridges must be protected from sunlight during storage;
•
Store the Check Cartridge at 2 - 32°C (36 -90°F) and humidity < 90%
•
Always store the Check Cartridge in its protective packaging to prevent
scratches which may affect the result.
•
If the Check Cartridge is refrigerated, allow to reach room temperature before
use. (~1 hour).
Precautions/Warings
•
For In Vitro Diagnostic Use.
•
Do not use the Check Cartridge beyond the expiration date.
•
Do not use the Check Cartridge if stored incorrectly or if it is dirty, scratched
or damaged.
Daily Check Cartridge
When to Use the Daily Check Cartridge
•
Once a day before samples are tested.
•
After moving the analyzer.
•
After an error message. (Er 1 or Er 3)
The Daily Check Cartridge consists of a cartridge without a reagent pack (packed in
a pouch).
Life-time: Until stated expiration date printed on the cartridge label.
17