13. SYSTEM CHECK
NOTE:
•
Store the Daily Check Cartridge in its
original package when not in use.
•
The Daily Check Cartridge can be
purchased separately from your local
distributor.
How to use the Daily Check Cartridge
®
1) Open the lid of the HemoCue
HbA1c 501 Analyzer.
2) Press
to enter Daily Check Cartridge mode.
3) Insert Daily Check Cartridge
while 'Daily Check' is
displayed.
18
13. SYSTEM CHECK
4) Close the lid. The test starts automatically.
5) After 1 minute 'OK' or 'Er 2' will be displayed.
6) After test completion, take out the cartridge.
NOTE:
•
If an error message is displayed, repeat the test.
•
Do not use the analyzer if the error persists.
•
To leave Daily Check mode, press
.
19