If one of the following conditions happens while the HV of the channel is On, the
NOTE
output of the channel is immediately set to Off, the LED on the front panel
indicates the corresponding Fault condition and the Fault information is sent to
the Serial output.
On "Remote" Connector on Rear Panel
"Fault" output: N.C. contact =
"HV 1,2,3" output: N.O. contact =
Ion Pumps Control Unit (Battery Backup) User Manual / 87-900-117-01 (D)
Change of the Local-Remote I/O switch status
Front panel HV switch set to Off (0)
External and/or HV Cable Interlock open
The contact is Open in normal operation (relay ON)
The contact is Closed in " Fault" condition
(relay OFF)
The contact is Open with HV Off (relay Off)
The contact is Closed with HV On (relay On)
Outputs relays rating: 1 A 48 Vac/dc
Installation Procedure
Error messages
4
75/94