Controller Description
Agilent's controller is an ion pumps controller.
It can drive up to 3 Ion Pumps simultaneously and independently.
The controller is designed to give the HV to the Ion Pumps when it is
connected to the Mains supply (Max. Output Voltage = 5000 Vdc) or it
is fed by a battery (Max. Output Voltage = 3000 Vdc). (See the
paragraph "Technical Informations" for details).
Agilent' s Ion Pump controller can be driven in Local/Remote I/O
operating mode or in the Serial mode via the RS232 port. The
selection of the operating mode is done by means of the
SERIAL/LOCAL-Remote I/O switch on the controller front panel.
In the LOCAL-REMOTE I/O mode, all the commands are given by the
front panel or by the connections to the DB25 "REMOTE" connector
on the rear panel.
The pump operating conditions data are anyway available on the
serial connection. In the SERIAL mode, all the commands are given
through the serial connection and all the pump operating conditions
data are available on the serial connection.
Storage
When transporting and storing the controller, the following
environmental requirements should not be exceeded:
Ion Pumps Control Unit (Battery Backup) User Manual / 87-900-117-01 (D)
temperature: from -20 ° to +70 °C
relative humidity: 0 – 95 % (non-condensing)
Installation Procedure