Herunterladen Diese Seite drucken

Agilent Technologies 829-5000R001 Bedienungshandbuch Seite 68

Werbung

4
Installation Procedure
Use
Tab. 1
REF.
SIGNAL
1
MAINS ON LED
2
HV1 ON/OFF
SWITCH
3
HV1 LED
4
HV2 ON/OFF
SWITCH
5
HV2 LED
6
HV3 ON/OFF
SWITCH
7
HV3 LED
8
SERIAL/LOCAL-REM
I/O SWITCH
9
BATTERY ENABLE/
DISABLE SWITCH
10
BATTERY LED
68/94
Ion Pumps Control Unit (Battery Backup) User Manual / 87-900-117-01 (D)
DESCRIPTION
When On it means that the unit is connected to the Power Line and the
Line Switch on the rear panel is in ON (I) position.
When in ON (I) position, it switches On the HV to the pump 1 if the
controller is set for LOCAL/REMOTE I/O operation and the external and
HV Cable interlocks are closed
It is On when the controller is supplying the HV to the pump 1.
It is flashing in case of Fault
(see paragraph "Technical Information" for details)
When in ON (I) position, it switches On the HV to the pump 2 if the
controller is set for LOCAL/REMOTE I/O operation and the external and
HV Cable interlocks are closed
It is On when the controller is supplying the HV to the pump 2.
It is flashing in case of Fault
(see paragraph "Technical Information" for details)
When in ON (I) position, it switches On the HV to the pump 3 if the
controller is set for LOCAL/REMOTE I/O operation and the external and
HV Cable interlocks are closed
It is On when the controller is supplying the HV to the pump 3.
It is flashing in case of Fault
(see paragraph "Technical Information" for details)
It allows to select the operating mode:
SERIAL: all the commands are given through the serial line
LOCAL-REM I/O: all the commands are given through the unit front
panel or through the 25 pin REMOTE connector on the rear panel
When in the ENABLE position, the battery is connected to the C.U. , in
case the Line Switch is OFF, or in case of a Power Failure, the 3 Ion
Pumps will be fed at 3000V generated by the Battery.
During the time the controller is connected to the Mains and the Line
Switch is ON, the battery is continuously recharged.
When in the DISABLE position, the battery is disconnected.
It is ON when the battery is enabled

Werbung