Installation Procedure
Error messages
Error messages
On Front Panel
During the controller operation, if an error condition is detected, the
HV LEDs on the front panel will give the indication:
HV 1,2,3
LEDS ON:
Led HV 1,2,3 flashing: error condition:
In Local-Remote I/O mode, the Reset of the Fault condition is
accomplished by switching to Off (0) the front panel switch of the
corresponding channel.
In Serial mode, the Reset of the Fault condition is accomplished by
sending the "F0n" command by the computer.
74/94
Ion Pumps Control Unit (Battery Backup) User Manual / 87-900-117-01 (D)
Normal Operation =
continuous flashing = overcurrent in Protect (I>8 mA)
1 flash every 2 seconds = Interlock HV cable
2 flashes every 2 seconds = remote Interlock
3 flashes every 2 seconds = HV Overoltage (Warning: do not switch
off HV)
4 flashes every 2 seconds = error inside ADC
5 flashes every 2 seconda = overcurrent on HV (short circuit)
(I>30mA for more than 30 seconds)
6 flashes every 2 seconds = the Local/Remote switch was moved
with HV ON
7 flashes every 2 seconds = overtemperature inside unit
HV ON