Instructions for carrying out inspection with service
On point 6:
Housing leakage currrent: The measurements are to be carried out in accor-
dance with DIN IEC 601,VDE 0750, Part 1/05.82, (Section 19.4 g) 5 and
must be carried out in conjunction with the pertinent power pack.
When using the plug power pack, the pertinent plug power pack can also be
measured on its own.
The leakage current between primary and secondary side of the plug power
pack corresponds with the maximum possible leakage current.
Plug power pack
Prim.
sec.
Isolation
transformer
HA
LN
AN
I NZ
Mains voltage +10%
1 <10hA
When using the integrated power pack:
The leakage current between primary and secondary side of the power pack
corresponds with the maximum possible leakage current.
isolation
transformer
I
Mains voltage +10%
1 <10pA