Faults:
PTA 004 Software
WARNING
FAULT
REMEDY
Red LED comes on
Operating indicator
at stilistand
Syringe flashes
1:4 Audible alarm
Staff call
Red LED comes on briefly
Operating indicator
at stillstand
Audible alarm
Red LED comes on briefly
Pressure in system
too high
Syringe empty
Syringe fitted
wrongly or not at all
Reset knob still released
Delivery rate not set
Display 00.0 !
Before rate change "C"
Key not actuated
Unit not started
Eliminate occlusion
Possible catheter
change necessary
Restart
Change syringe
Restart or end in-
fusion
Switch off unit
Fit syringe correctly
Restart
Engage reset knob
Restart
Enter delivery rate
Restart
Enter delivery rate
Operating indicator
Restart
at stilistand
Audible alarm
Start unit
Staff call
Syringe flashes
Prealarm 3 min.
System is not faulty
1:9 Audible alarm
Operating indicator
before end of infusion
Since unit is running
normally and giving
rotates
prealarm signal
Accumulatorlogo
Accumulator prealarm
Continue to run unit
hes
1:9 Audible alarm
Operating indicator rotates
Battery flashes
1:4 Audible alarm
Operating indicator
at stillstand
Red LED flashes
Staff call
Red LED comes on
Operating indicator
at stillstand
Continous tone
Staff call
Red LED comes on
Continous tone
Operating indicator
shows 9 spokes
Staff call
Red LED comes on
Continous lone
Staff call
1:4 = 1 sec. audible alarm
4 sec. break
1:9 = I sec. audible alarm
9 sec. break
Accumulator empty
Accumulator prealarm
signal has been over-
locked
Accumulator empty
Unit has broken down
During selfcheck at
keyboard button was
pressed
from mains
Switch off immediately
or operate from mains
Swich
off
immediately
and charge accumulator
Send unit for repair
Switch off the unit
and start again
Ovese Service-Untertage dart ohne unsere Genenmigung
uch auszuaswerse weoer verweltatgt noch Dritten zu-
whole, nor made accessible to Third partes without our
consent, and remains the property Of our company.
pangiich gemacht werden und Dierbt mit allen Rechten
Page
Edition
B Braun Melsungen È
4
|
©
unser Eigentum.
This maintenance reterence must not De copied, partiyor
Perf. sec.FT
[9 1 ]
4
09 /9 1
TECH
; |
|
e
E

Werbung

loading