Herunterladen Diese Seite drucken

B. Braun PERFUSOR-secura FT Technische Beschreibung Und Bedienungsanleitung Seite 51

Werbung

Special functions with test adapter
Perfusor-secura FT
Software Version PTA 004
Software version display
Programming of CC number (Ident. No.)
Calling up the working hours
Calling up accumulator threshold
Adjusting the accumulator threshold
The threshold value will be raised by 1 by actuating the comma key when calling up the ac-
cumulator threshold value. (b00...b15 overflowing to 0)
Automatic compensation of accumulator threshold
Set accumulator threshold manually to 0. Then charge the unit for 20 hours and discharge for
6 hours in accumulator operation. The accumulator threshold is now balanced out automatically,
call up accumulator threshold value once again and check whether the accumulator threshold has
been raised to a value approx. 0; if the accumulator threshold value has not been raised then the
accumulator is defective or the balancing out instructions were not carried out correctly.
Important
Alternatively the value can also be taken over and given in from the first circuit board.
The rules of CMOS Technic have to be noted when wor-
king on circuit boards!