Faults
Warning
Fault
Remedy
Red LED comes on
Oper. indicator at standstill
Syringes flashes
Audible alarm 1:4
Staff call
Pressure in system too high
Eliminate occlusion
Possibly catheter change
necessary
Restart
Syringe empty
Change syringe
Restart or end infusion
Switch off unit
Syringe fitted wrongly
Fit Syringe correctly
Audible alarm
Red LED comes on briefly
Oper. indicator at standstill
Audible alarm
C-key not activated before
changing delivery rate
or not at all
Restart
Reset knob still released
Engage reset knob
Restart
Red LED comes on briefly
Delivery rate not set
Enter delivery rate
Oper. indicator at standstill
Display 00.0!
Restart
Press C-key, enter rate
Restart
Unit has not startet
Enter delivery rate
Restart
Start unit
Syringe flashes
Audible signal 1:9
operating display rotates
Pre-alarm 3 min before
end of infusion
No malfunction since
unit is running normally and
giving pre-alarm signal
Accumulator flashes
Audible signal 1:9
operating indicator rotates
Accumulator pre-alarm
Continue to run unit from
mains
Battery flashes
Audible alarm 1:4
Oper. indicator at standstill
Red LED flashes
Staff call
Accumulator discharged
Accumulator pre-alarm signal
has been overlocked
Switch off immediately or
operate from mains
Red LED comes on
Oper. indicator at standstill
Continuous tone
Staff call
Accumulator empty
Switch off immediately
change accumulator
Red LED comes on
Oper. indicator at standstill
All symbols signal
Continuous tone
Staff call
Technical failure
Send unit for repair
indicator shows 9 spokes
Continuous tone
Red LED comes on, operating
During selfcheck a keyboard
button was pressed
Switch off the unit and start
again
1:4
1:9
20
= 1 s audible alarm, 4 s break
= 1s audible alarm, 9 s break