6.2 CRYO PROBE CONNECTION
Observe sterilisation protocol before using a Cryo Probe (see Section 3). Allow Cryo Probe to
cool to room temperature after a sterilisation.
Before connecting the Cryo Probe inspect it for signs of obvious damage.
1. Remove sterilisation cap from the Probe.
2. Connect the Cryo Probe to the Console by inserting it into the
coupling and pushing against the spring-operated collar until a
positive click is heard. When the Probe is properly connected the
relevant Probe Symbol is displayed together with the Run option.
3. To proceed, the Run button
4. If a Reusable Probe has been connected, the system automatically initiates a purging
cycle of 90 seconds.
During the purge cycle an animated Wait Symbol is displayed alongside the Probe
Symbol. Three short beeps signal that the purge cycle is complete.
5. The equipment is now ready for use as indicated by the timer and the Ready Symbol
6. If a Disposable Probe has been connected, there is no purge cycle.
While the Probe is purging, all Footswitch operations are disallowed to ensure that the Cryo
Probe has completed the minimum purge.
6.3 FREEZE / DEFROST CYCLES
WARNING: Probe tip reaches extremely low temperature in use
(between -20ºC and -60ºC depending on Probe type used).
Only the areas of the eye to be treated should make contact with
the probe tip or probe shaft.
Freezing of the Cryo Probe is controlled manually by the operator using the Footswitch.
EN
must be pushed.
16
.