Herunterladen Inhalt Inhalt Diese Seite drucken

Intended Use; Safety Installation On The Device; Set-Up; Unpacking - Heidolph MR Hei-End Betriebsanleitung

Sicherheitsmagnetrührer
Vorschau ausblenden Andere Handbücher für MR Hei-End:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
The MR Hei-End is a safety magnetic stirrer with a heating function, suitable for non-
monitored operation. The Hei-End is a magnetic stirrer featuring a hotplate function. It is
designed for use in chemical and biological laboratories of industrial enterprises,
universities, schools and pharmacies.
The instrument is designed to heat substances to a specified temperature and/or mix them
through the action of a magnetically linked stirring rod in a vessel.
To ensure maximum service life, observe the specified ambient conditions (temperature
and humidity) and ensure that the instrument is not exposed to a corrosive atmosphere.
Function of the safety switching-off facility:
2 independent, temperature-measuring sensors are located in the hotplate. One serves
the processor-driven control of the hotplate. The other serves the analog safety circuit
switching off all the poles.
Criteria for the safety switching-off function (heating OFF):
 The temperature set for the hotplate (reference value) is exceeded by the set safety
temperature (max. 25°C) (see chapter 9.2).
 The measured temperature of the hotplate (actual value) is higher than 360°C
 Sensor breakage or short circuit in the temperature sensor of the safety circuit
 Sensor breakage or short circuit in the temperature sensor of the control circuit
 The µ-controller is affected in the program sequence
 Sensor breakage or short circuit in in saftey circuit of external temperaure sensor
 Sensor breakage or short circuit in in control circuit of external temperaure sensor
1

Unpacking

Unpack the instrument with care. Inspect for damage and report such damage or missing
parts to your supplier right away.
2

Setting up the instrument

Place the instrument on a solid, horizontal surface. Keep the area around the instrument
clear of other items.

INTENDED USE

SAFETY INSTALLATION ON THE DEVICE

SET-UP

29
E

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis