Inserting a Specimen Holder
Adjusting the Force
Adjusting the Preparation Time
Adjusting the Position of
RotoForce-3
Running the preparation
RotoForce-3
Instruction Manual
Quick Reference Card
n Place the specimen holder under the quick coupling.
n Press and hold the release button on the column of RotoForce-
3, while guiding the tap of the specimen holder into the
coupling.
n Turn the specimen holder until the three pins engage in the
corresponding holes.
n Release the button.
fO
n Press FORCE
value.
tO
n Press TIME
value.
n Unlock RotoForce-3 by moving the handle up.
n Adjust the RotoForce-3 so that the specimens are running at
about 2 - 4 mm from the periphery of the preparation disc.
n Lock RotoForce-3 in that position by moving the handle down.
n Adjust time and force on the RotoForce-3.
n Set RotoForce-3 to Auto.
n Set Lupo to Auto if lubricant dosing is required.
s
n Press START
fN
or
to adjust the force to the desired
tN
or
to adjust the time to the desired
on RotoPol.