Connecting RotoForce-3
Electrical Connection
Connection to Compressed Air
Connection to
the Preparation Equipment
4
RotoForce-3
Instruction Manual
RotoForce-3 is supplied with electric cable, air hose and
connection cable.
Mount a plug on the cable or connect directly to mains:
Black and brown: phase
Yellow/green: earth
Check that the mains voltage corresponds to the voltage stated on the type
plate on the back of the machine.
Check that the specimen holder rotates anti-clockwise.
n Mount an air hose on the enclosed quick coupling and secure
with the enclosed hose clamp.
n Connect the quick coupling to the factory-mounted air hose of
RotoForce-3.
The air pressure must be between 5 bar (73 psi) and 6.5 bar (95 psi)
The air must not contain water, oil or particles. If necessary, a pressure
regulator/filter unit must be installed between air supply and RotoForce.
RotoPol alone: Connect the communication cable to the back of
RotoPol (see the Instruction Manual of RotoPol).
RotoPol with Multidoser and/or RotoCom: Connect the
communication cable to the back of Multidoser (see the
Instruction Manual of Multidoser)
IMPORTANT
IMPORTANT