Weight
Material
Terminals
Display elements
Operating elements
Remote operation
• Approx. 1.6 kg (3.53 lb) (aluminium housing)
• Approx. 4.2 kg (9.26 lb) (stainless steel housing)
Housing
Die-cast aluminum housing AlSi10Mg with powder coating
on polyester basis
Stainless steel 1.4435 (AISI 316L)
Cables / wires up to max. 2.5 mm
12.0.8
Human interface
Fig. 17:
LC display of the field indicator (illuminated, pluggable in 90° stages)
Item 1: bargraph display in increments of 10% with indicators for measuring range undershoot (item 1a)/overshoot (item
1b)
Item 2: warning sign in the case of limit value violation
Item 3: unit indicator K, °F, °C or %
Item 4: measured value display (character height 20.5 mm/0.8")
Item 5: status and information indicator / configuration
Item 6: 'programming disabled' indicator
• Display range
-19999 to +99999
• Offset
-19999 to +99999
• Signalling
Measuring range overshoot/undershoot
• Limit value violation
Lower/upper limit value exceeded
3-key operation (-/+/E) integrated in device, access with housing open
Configuration
The device is configured with the ReadWin
Interface
Configuration interface at device; connection to PC via configuration kit (see "Accessories").
Configurable device parameters (selection)
Measuring dimension, measuring ranges (linear/square), setup block using user code, failsafe mode,
digital filter (damping), offset, limit value (min/max/alarm), alarm limit values freely adjustable.
Nameplate
Aluminum AlMgl, anodized in black
1.4301 (AISI 304)
2
(AWG 13) plus ferrule
®
2000 PC operating software.
Technical data