Cleaning Recommendations for the
Sterilizable Torque Base
Notice
Product damage will result if you do not follow these
notices when cleaning the sterilizable torque base:
•
Do not use ultrasonic or high pressure washers.
•
Do not use chlorinated substances such as
bleach solution.
•
Do not clean with abrasives.
1. Remove all gross matter (blood, mucous, and
tissue):
a. Dampen a soft cloth with a cleaning agent.
b. Thoroughly wipe all surfaces of the base.
Follow the procedures approved by your
institution.
2. For at least two minutes, soak in an enzymatic
cleaning agent (for example, Klenzyme® or Enzol®)
according to manufacturer's instructions.
3. Scrub all surfaces with a small soft brush.
4. Rinse thoroughly with water.
5. Dry with a clean, soft cloth.
Sterilization Recommendations
1. Load the sterilizable torque base into a CUSA
Excel sterilizer tray.
2. Steam sterilize the base using procedures
approved by your institution.
Refer to the CUSA Excel System User's Guide for
the steam sterilization parameters.
Refer to the CUSA Excel user manual for more
information.
Any serious incident that has occurred in relation
to the device for the user and/or the patient should
be reported to the manufacturer and the competent
authority of the member state in which the user and/or
patient is established.