TROUBLESHOOTING
If lamp is not
illuminating and . . .
fan is not operating.
fan is operating.
Light Source is not plugged
into either the wall outlet or
the power cord socket.
Power switch is not
turned on.
The lamp door is not
properly
installed.
One or both fuses have
blown.
Power cord is damaged.
This may cause intermittent
operation.
No power is supplied to
the wall outlet.
The lamp was not able to
"cool down" before restarting.
The Light Source has
overheated.
The lamp has burned out.
11
Plug the power cord into the
wall outlet and power cord
socket at back of unit.
Depress the "I" side of the
power switch.
Remove the lamp door and
reinstall carefully, making
sure that the lip on the
bottom fits into the front of
the housing and that the
thumbscrew is fully tightened.
Replace both fuses. Follow
the instructions in the fuse
replacement section. If fuses
continue to blow, return Light
Source to an authorized
repair center.
Replace the power cord with
a new power cord.
Verify that the wall outlet
has power by checking the
circuit breaker status and by
operating
other
equipment
at that wall outlet.
Wait at least 15 seconds
after turning the Light Source
off before turnina it back on.
Verify that all ventilation
slots and the fan area are not
obstructed. The Light Source
should always have
at least 3-inches clearance.
Replace with a new SdlarcTM
lamp. Refer to the lamp
replacement
section.