NOTE: Replace with Stryker lamp #502 220 176 only. Do not touch the
lamp itself, or the interior reflective surface of the lamp. Skin oils on the
lamp will cause premature lamp failure. Always hold lamp by the ceramic
connector base. Remove any grease or fingerprints with a clean cotton
swab moistened with alcohol. Do not leave any lint on the lamp.
5. To install a new lamp, grasp the lamp by the ceramic base and be
sure that the alignment pin on the lamp is pointing up. Gently slide the
lamp down into place between the holding spring assembly and the front
of the box, allowing the holding spring to help seat the lamp in the fixture.
Check to be sure that the
alignment pin on the lamp is
securely placed in the slot on
the fixture. When the lamp is
properly seated, you will be
able to rotate the lamp slight-
ly in either direction. If lamp
does not rotate at all, or
rotates freely, pull back gen-
tly and reposition lamp as
described above. Plug the
black connector assembly
wire back into the lamp.
6. Reattach the lamp door securely. The unit will not power up unless the
lamp door is seated properly.
assembly
7
Be sure
alignment pin
and lamp are
properly seated
black connector
'wire