g) AC voltage measurement
Range
600.0 mV
6.000 V
60.00 V
600.0 V
LoZ ACV 600.0 V
LPF ACV 600.0 V
Input impedance: approx. 10 MΩ.
True RMS display.
Frequency response: 40 - 400 Hz. LPF frequency response: 40 - 200 Hz.
After using the LoZ function, please cool down the meter for 1 minute.
Accuracy guarantee range: 5~100% of range, short circuit allows least significant digit
<5.
The AC crest factor is ≤2.5 when measured at 4000 counts. The full range AC crest factor
of 6000 counts is ≤1.8.
Non-sinusoidal waveforms:
- When the crest factor is 1.0 to 2.0, the accuracy must be increased by 4.0 %.
- When the crest factor is 2.0 to 2.5, the accuracy must be increased by 5.0 %.
- When the crest factor is 2.5 to 3.0, the accuracy must be increased by 7.0 %.
Input voltage: max. 600 Vrms.
68
Resolution
0.1m V
0.001 V
0.01 V
0.1 V
0.1 V
0.1 V
Accuracy
± (1.5 % + 5)
± (1.3 % + 4)
± (1.3 % + 4)
± (1.3 % + 4)
± (2.6 % + 4)
± (2.5 % + 6)