m) AC current measurement
Range
600.0μA
μA
6000μA
60.00 mA
mA
600.0 mA
6.000 A
A
10.00 A
When the measured current is >5 A, each measurement time should be ≤30 s and the
rest interval should be ≥15 minutes.
True RMS display.
Frequency response: 40 - 400 Hz.
Accuracy guarantee range: 5 - 100 % of range, open circuit allows least significant digit
<5.
The AC crest factor is ≤2.5 when measured at 4000 counts. The full range AC crest factor
of 6000 counts is ≤1.8.
Non-sinusoidal waveforms:
- When the crest factor is 1.0 - 2.0, the accuracy must be increased by 4.0 %.
- When the crest factor is 2.0 - 2.5, the accuracy must be increased by 5.0%.
- When the crest factor is 2.5 - 3.0, the accuracy must be increased by 7.0%.
Overload protection:
- F1 Fuse: Φ6 x 32 mm, FF 10 A, H 600 V, Breaking capacity:10 KA
- F2 Fuse: Φ5 x 20 mm, FF 600 mA, H 600 V, Breaking capacity: 500 A min.
Resolution
Accuracy
0.1 μA
1 μA
± (1.3 % + 4)
10 μA
0.1 mA
1 mA
± (1.6 % + 4)
10 mA
± (1.8 % + 6)
71