8. How to recall 30 readings in Memory Mode

This thermometer can recall the last 30 readings.
 Recall mode AP: Press the START button 3 to enter
Recall mode when power is off. The memory icon «M»
will flash.
 Reading 1 - the last reading AQ: Press and release the
START button 3 to recall the last reading. Display 1
alone with memory icon.
 Reading 30 - readings in succession: Press and
release the START button 3 consecutively to recall the
last 30 readings in succession.
Pressing and releasing the START button 3 after the last
30 readings have been recalled will resume the above
sequence from reading 1.

9. Error Messages

 Measured temperature too high AR: Displays «H»
when measured temperature is higher than 42.2 °C /
108.0 °F in body mode or 100 °C / 212 °F in object mode.
 Measured temperature too low AS: Displays «L»
when measured temperature is lower than 34.0 °C /
93.2 °F in body mode or 0 °C / 32 °F in object mode.
5
 Ambient temperature too high BT: Displays «H» in
conjunction with the
when ambient temperature is
higher than 40.0 °C / 104.0 °F.
 Ambient temperature too low BK: Display «L» in
conjunction with the
when ambient temperature is
lower than 16.0 °C / 60.8 °F in body mode or lower than
5.0 °C / 41.0 °F in object mode.
 Error function display BL: The system has a malfunction.
 Blank display BM: Check if the battery has been loaded
correctly. Also check polarity (<+> and <->) of the battery.
 Flat battery indication BN: If the steady battery icon is
the only symbol shown on the display, the battery should
be replaced immediately.

10.Cleaning and Disinfecting

Use an alcohol swab or cotton tissue moistened with alcohol
(70% Isopropyl) to clean the thermometer casing and the
measuring probe. Ensure that no liquid enters the interior of
the thermometer. Never use abrasive cleaning agents, thin-
ners or benzene for cleaning and never immerse the instru-
ment in water or other cleaning liquids. Take care not to
scratch the surface of the probe lens and the display.

11.Battery Replacement

This instrument is supplied with one lithium battery, type
CR2032. Replace with a new CR2032 battery when the
flashing battery symbol appears on the display BN.
Remove the battery cover by sliding it in the direction
shown. Remove the battery and replace with a new one BO.
Batteries and electronic instruments must be dis-
posed of in accordance with the locally applicable
regulations, not with domestic waste.

12.Guarantee

This instrument is covered by a 2 year guarantee from the
date of purchase. The guarantee is valid only on presenta-
tion of the guarantee card completed by the dealer (see
back) confirming date of purchase or the receipt.
 The guarantee covers the instrument. Batteries and
packaging are not included.
 Opening or altering the instrument invalidates the guar-
antee.
 The guarantee does not cover damage caused by
improper handling, discharged batteries, accidents or
non-compliance with the operating instructions.
Please contact Microlife-service.