13.Technical Specifications

Type:
Non Contact Thermometer NC 120
Measure-
Body Mode: 34.0-42.2 °C (93.2-108.0 °F)
ment range:
Object Mode: 0-100.0 °C (32-212.0 °F)
Resolution: 0.1 °C / °F
Measure-
Laboratory:
0.2 °C, 36.0  39.0 °C
ment accu-
(0.4 °F, 96.8  102.2 °F)
racy:
Display:
Liquid Crystal Display, 4 digits plus special
icons
Acoustic:
The unit is turned ON and ready for the
measurement: 1 short beep
Complete the measurement:
1 long beep (1 sec.) if the reading is less
than 37.5 °C (99.5 °F), 10 short «beep»
sounds, if the reading is equal to or greater
than 37.5 °C (99.5 °F).
System error or malfunction:
3 short «bi» sounds
Memory:
Auto-Display the last measured temperature
30 readings recall in the Memory Mode
NC 120
Backlight:
The display light will be GREEN for
4 seconds, when the unit is turned ON.
The display light will be GREEN for
5 seconds, when a measurement is
completed with a reading less than 37.5 °C
(99.5 °F).
The display light will be RED for 5 seconds,
when a measurement is completed with a
reading equal to or higher than 37.5 °C
(99.5 °F).
Operating
Body Mode: 16-40.0 °C (60.8-104.0 °F)
temperature:
Object Mode: 5-40.0 °C (41-104.0 °F)
Storage
-20 °C to +50 °C (-4 °F to 122 °F)
temperature:
15-95 % relative maximum humidity
Automatic
Approx. 1 minute after last measurement
Switch-off:
has been taken.
Battery:
CR2032 Battery (X1) 3V - at least 1000
measurements
Dimensions: 120 x 28 x 21 mm
Weight:
39 g (with battery), 36 g (w/o battery)
Reference to
ASTM E1965;
standards:
IEC 60601-1; IEC 60601-1-2 (EMC)
This device complies with the requirements of the Medical
Device Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial tech-
nical inspection is recommended for professional users.