10. Faults and cures

Fault messages are indicated on the display as „E"
followed by a four digit number.
Each number indicates a particular fault condition:
- „0" = no fault
- „1" = fault is present
Display
Exxx1
There are no valid
values in the
memory (RAM).
This condition
could be caused by
changing the
battery.
Exx1x
No valid operating
parameters in the
EEPROM.
Ex1xx
No valid calibration
values in the
EEPROM.
E1xxx
CPU error, This
means fault in the
electronics.
Cause
Loss of stored
measured values,
Counter values as
well as date and
time.
Loss of operating
settings.
Unit not ready for
operation.
Unit defective.
10. Faults and cures
Effect
Set the unit up
again using
ReadWin.
Set up the unit
again using
ReadWin .
Please return the
unit for repair.
Please return the
unit for repair.