6.4 Changing the battery
If the battery charge goes below the minimum capacity required
during a measurement cycle the low battery indicator is
displayed.
In this case measurement continues for approx. 30 days. In
order to achieve continuous recording the data should be read
out and the battery changed as swiftly as possible.
Changing the battery:
Attention: First read out all stored data from the unit using
ReadWin.
Open the unit cover and hinge downwards.
Switch the unit off using the on/off switch
Check that the unit is switched off by monitoring the display.
Push the opening levers on the battery terminals
pull the battery out of the holding clamp.
Attention: In order to avoid data loss connect a new battery
(order No. 50085928) into the unit within 5 min.:
Bend battery connection wires to the required angle.
Attention: Note battery polarity.
The blue terminal is the minus pole and the orange terminal is
the plus pole.
Push the opening levers on the battery terminals
push in the battery connection wires.
Push the battery into the battery holder.
Switch the unit on using the on/off switch
Check the unit function by monitoring the display.
Close the unit cover.
Set up date and time using ReadWin.
If the battery is not changed within 30 days of the low battery
display being active the unit changes into a stand-by mode.
This means the display is turned off and data is no longer
recorded.