6.1.2 Testing the sensor with the METTLER TOLEDO
O
Sensor - Master InPro 6800
2
We recommend to use the METTLER TOLEDO O
sor-Master InPro 6800 to check the quality of your sen-
sor as follows (not applicable for digital ISM sensors):
• Connect the sensor to the O
As soon as the sensor is connected to the O
Sensor-Master, the polarization function is auto-
matically activated. Please note: if the sensor was
disconnected from the transmitter for longer than 5
minutes, the sensor must be polarized first (polar-
izing time see «Section 5.1») to get representative
test results.

• Battery Check:

Push the switch to the left. If the battery is ok and
the O
lights up. Otherwise, please consult the instruction
manual of the O
• Sensor Check:
For this test the O
and the membrane of the sensor must be dry
and clean.
Expose the sensor (connected to O
ter) to the air. By pushing the switch to the right
to the position "2 – Sensor check", the O
Master checks wether the electrode current for air
measurement delivered by the sensor is within the
admissible range (40 to 100 nA for InPro 6800). If
the green LED lights up the current for air measure-
ment is within the admissible range.
InPro 6800
52 200 953 C
Sensor-Master is operational the green LED
2
2
InPro 6800 O
Sensor-Master.
2
LED
Switch
Sensor-Master.
sensor must be fully polarized
2
© 02 / 2023 METTLER TOLEDO
Sensor 12/25 mm
2
Sensor- Mas-
2
2
Printed in Switzerland
Sen-
2
2
Sensor-