Display
mV/V of LC1 1.99995
Load (mV/V): 1.40001
Information
You can use the Clr key to clear the value and then re-enter it.
Press the F2 key 10 times to enter the decimal point for comma values. The decimal
separator then appears in the display.
8.8
Linearizing the scale
Linearization compensates for any deviation of the scale characteristic curve from the
ideal straight line by setting correction points that correspond to the actual course of the
scale curve.
The linearization software supports up to 6 correction points, though in practice 1 or 2
are usually enough. The linearization points should be placed at the points on the weight
curve where the greatest variations occur. Linearization points must be lower than the
set maximum load. The sections between adjacent linearization points are calculated as
straight lines, meaning the values between them are interpolated in linear mode.
Display
Standard
WTX110 V#.#.#
Password
1 Service Mode
WTX110
CALIBRATION AND ADJUSTMENT
Key
Explanation
Enter the sensitivity of the first load cell.
Continue with the next load cell. After the last
load cell continue with the next step.
Display of the signal for the maximum load
(2100 kg).
Enter the (averaged) signal for the maximum load
and confirm with F5.
Key
Explanation
Prepare to switch to Service Mode; the WTX
version is displayed.
Display of the current firmware version.
Press within 3 seconds to access the password
input.
Enter the password 324.
Call Service Mode.
93