Display
Unit
8.4

Geo value setting

Before calibrating, set the calibration location (e.g. Germany = 20) using the geo value
table below. This allows you to calibrate a scale in one location and then set it up in
another. Once there, all you have to do is enter the geo value of the installation site,
because of the different gravity. The scale does not then need to be recalibrated.
To access the input area select Service Mode ↓ Interface → General →
Calibration ↓ Calibrate Scale 1 ↓ Select Group 1 - 9 → Scale
Parameters → Calibration ↓ Geo value. See section 6.1, page 49 on how to
access Service Mode, and
Information
For details on entering numbers via the WTX110 display see section 5.2.2
"Entering whole numbers", page 47.
Country
France
Finland
Belgium
Denmark
Germany
UK
Ireland
Norway
Netherlands
Austria
80
Key
Explanation
Enter the unit:
kg
kg
g
t
lb
N
Back to Select Group.
Navigation
Pilot.
kilogram
gram
metric ton
pound
Newton
Geo value
20
24
21
23
20
21
22
24
21
19
CALIBRATION AND ADJUSTMENT