The resolution of the analog output signal is 15 bits (32,768 increments). The output
signal is active and potential-free.
In Service Mode choose Interface → Config. → Config. Scale → Config.
Analog Out → DAU15 to specify whether the analog output is to output voltage or
current: 0/2 ... 10 V or 0/4 ... 20 mA.
Information
You must always adjust the DAU15 module (AO option).
Fig. 4.12
Connection of analog output to KL3
KL3: Analog output DAU15 in slot PIM1
1
2
3
I+
4
I‒
5
6
U+
7
UI‒
8
30
Current output 0/4 ... 20 mA (+)
Current output 0/4 ... 20 mA (–)
Voltage output 0/2 ... 10 V (+)
Voltage output 0/2 ... 10 V (–)
INSTALLATION AND ELECTRICAL CONNECTION
WTX110