Adapting Scale Parameters - HBK WTX110 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Display
Save parameters?
" There are linearization points
Fixpoints active!
8.10

Adapting scale parameters

To adapt to the weighing environment, set the following parameters so as to obtain an
optimum weighing result. Observe the restrictions for legal-for-trade scales; see section
8.10.1. The procedure and the exact meanings of the settings are explained in section
8.10.2, page 98.
Menu item
Motion window
Motion Counter
Filter size
Auto zero range
Pushbutton zero
Power up zero
Overload
NTEP approval?
Underload 20d?
Update rate
WTX110
CALIBRATION AND ADJUSTMENT
Key
Explanation
N
Y
N
Confirm selection.
Linearization entries have already been made.
They can influence the calibration if the new zero
point differs from the one measured during
calibration. Linearization should always be
performed after a complete calibration.
Explanation
Size of the motion window for standstill monitoring.
Number of weight values for standstill monitoring
(motion counter).
Filtration level of the digital filter (with an unstabilized weight
indicator).
Size of the range of zero setting for automatic zero setting.
Size of the range of zero setting for the Zero key.
Size of the range of zero setting on start-up.
Weight limit for display hide-on-overload function.
Metrological properties according to American requirements.
Hide weight indicator if gross weight values less than -20d.
Select the update rate for the digital filter.
Save the new zero value
Discard all changes
97