Error message
Maximum 20 steps!
Method database is full!
Method group is full!
Cleaning not allowed during dosing
process!
Suspension and lubricant not
compatible!
Surface and suspension not
compatible!
Dosing denied, the following
suspension is not configured:
DP-P 9 μm (Example)
Dosing denied, the following
lubricant is not configured:
Blue (Example)
The method is used for process,
certain functions are not allowed!
Process in progress certain
functions are not allowed!
TegraDoser-5
Instruction Manual

6. Trouble-shooting

Explanation
Each method can contain a
maximum of 20 steps.
Database for User Methods is full.
Each group can contain a maximum
of 20 methods.
It is not possible to perform cleaning
while a process is running .
The selected User Lubricant does
not have the correct "Lub. type" or
cannot be used together with the
selected Suspension.
For Example:
The selected Suspension cannot be
used with water-based Lubricants.
The selected Suspension does not
have the correct "Abr. type" or
cannot be used together with the
selected Surface.
For Example:
The selected Surface can only be
used with OP- Suspensions and the
selected Suspension contains
diamonds.
The preparation process cannot
start, as the Suspension required is
not configured into the
TegraDoser-5.
The preparation process cannot
start, as the Lubricant required is
not configured into the
TegraDoser-5.
The attempted function cannot be
performed as the method is
currently being used for a process
For Example:
Methods cannot be deleted when
they are being used to run a
process.
The attempted function cannot be
performed while a process is
running.
For Example:
Bottle Configuration cannot be
changed when a process is running.
Action required
Delete any User Methods that
are no longer required.
Delete an existing method
no longer required
.
Clean the tubes after the
is completed.
Change the "Lub. type" or choose
another Lubricant.
Change the "Abr. type" or choose
another Suspension.
Enter the Bottle Configuration menu
and set up the required Suspension.
For Example: DP-P 9 μm.
Enter the Bottle Configuration menu
and set up the required Lubricant.
For Example: Blue.
Wait until the process is
completed before performing the
function.
Wait until the process is
completed before performing the
function.
that is
process
59