Maintenance

4.3 Cleaning the interior

The load plate has to be removed in order to clean the load cell.
NOTICE
Damage to the load cell due to improper handling.
▲ Never touch, direct compressed air against or spray the rubber membrane (1) resp.
the rubber cover (2) of the load cell.
1. Remove load plate, for PBD7..-CC / PBA7..-CC see section 4.5.
2. Blow out dirt or flush out with a medium-power water jet (< 2 bar).
3. Replace load plate, for PBD7..-CC / PBA7..-CC see section 4.5.

4.4 Subsequent treatment

In order to protect the scale, carry out the following subsequent treatment:
➜ Rinse the scale with clear water and remove cleaning agent completely.
➜ Dry off the scale with a lint-free cloth.
➜ With stainless steel scales, treat the inside and outside with an oil suitable for
foodstuffs.
PBD7- / PBA7-series 30403592A