FIRMNESS CATEGORIES
EN
Refer to the chart below to determine the correct firmness category.
Note: Incorrect category selection may result in poor device function. Contact College Park Technical Service if
you have questions about category selection.
WEIGHT LBS
WEIGHT KG
SIZE CM
LOW IMPACT
ASSEMBLY AND DISASSEMBLY (FOR SOCK REPLACEMENT)
Use the Foot Horn to don and doff the foot shell. Remove the CPI sock and replace as needed.
ENDOSKELETAL MOUNTING
Use only high quality proximal endoskeletal components.
STATIC ALIGNMENT
For optimal function, balance the patient's weight evenly between the heel and toe (Figure 4).
▪ The Celsus iVT was designed with a 3/8" (10.0 mm) heel rise.
▪ The load line divides the foot at 1/3 heel lever and 2/3 toe lever.
DYNAMIC ADJUSTMENTS
Desired Result
Firmer Toe Response
Softer Toe Response
Firmer Heel Response
Softer Heel Response
INSTALLING THE CELSUS IVT
College Park recommends the iVT valve kit to connect the system to the socket. Use of other products is the
sole responsibility of the provider.

TUBE CONNECTION

1.
Attach the tube to the hose connector.
2.
Cut the tube to a proper length and connect to the socket.
3.
Use the supplied tube retainer to secure the tube to the pylon or other adapter.
4.
Position the tube on the medial side to avoid damage.
5.
Once assembled verify the pump is working properly; creating vacuum during gait and not leaking air.
TUBE INSPECTION
The tube is the most sensitive part of the Celsus iVT and should be fitted carefully. Verify the tube is NOT:
▪ Folded or pinched (Figure 2)
▪ Contacting sharp objects
▪ Punctured
CHANGING HOSE CONNECTOR SIDE
The hose connector is pre-assembled for a left foot and can be reconfigured for a right foot.
1.
Use a 5/16 wrench to loosen the hose connector and exhaust vent.
2.
Switch the positions. Ensure the check valves are oriented in the correct direction for proper function (Figure 3).
3.
Retighten. Torque to 35 in-lbs (4 N-m).
0-140
141-180
0-63
64-81
21-30
1
2
Alignment Change
Plantarflex the Celsus iVT or move load line posterior
Dorsiflex the Celsus iVT or move load line anterior
Dorsiflex the Celsus iVT or move load line anterior
Plantarflex the Celsus iVT or move load line posterior
▪ Dangling or hanging during gait
▪ Located laterally, to avoid damage
181-220
221-300
82-100
101-136
25-30
3
4