CMA
RANGE1 / RANGE 2
If a voltage of 0 V is present at PIN RANGE1/RANGE2, measuring range 1
(100% F
) is active at the charge amplifier. The charge amplifier gives you
nom
the option to zoom into a second measuring range (20% of the nominal
(rated) force / output span). A voltage of 24 V
RANGE1/RANGE for this.
10 V
2 V
0 V
0%
RANGE 1 / RESET
Fig. 6.3
Zooming into the second measuring range
Changing the measuring range to 20% can also be used to operate the force
transducer with high safety reserves for measurement tasks where overload is
critical. The charge amplifier overloads above 20%.
A2559−1.0 en/de/fr
10 V
0 V
0%
100%
must be applied at PIN
DC
20%
RANGE 2 / RESET
15
100%