Herunterladen Inhalt Inhalt Diese Seite drucken

Interconnections; P3 - Vent; P5 - External Fan; J5 - In-Out - Agilent Technologies TV 301-P Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9

INTERCONNECTIONS

The following figure shows the TV 301-P interconnections.
SERIAL LINE
MAINS
EXTERNAL
FUSE TYPE
250V T 8A
FAN
VENT VALVE
LD1
STATUS
P3
P5
J6 SERIAL
VENT
ADDRESS
FAN
ADDRESS
STATUS
J2 PROFIBUS
MSD
LSD
STATUS

P3 - Vent

This is a dedicated 24 Vdc connector to control the
optional vent valve.
P5 – External fan
This is a dedicated 24 Vdc connector to supply the
optional external fan.
MAINS POWER
IMPUT/OUPUT LINE
TO PUMP
PROFIBUS LINE
J5 IN-OUT

J5 - IN-OUT

This connector carries all the input and output signals to
remote control the TV 301-P Navigator Controller.
It is a 15-pins D type connector; the available signals
are detailed in the table, the following paragraphs de-
scribe the signal characteristics and use.
PIN N.
1
START/STOP (+)
2
START/STOP (-)
3
INTERLOCK (+)
4
INTERLOCK (-)
5
SPEED SETTING (+)
6
SPEED SETTING (-)
7
SOFT START (+)
8
SOFT START (-)
9
+24 Vdc
10
SPARE
11
PROGRAMMABLE SET POINT
12
SPARE
13
FAULT OUTPUT
14
PROGRAMMABLE ANALOG
SIGNAL (+)
15
GROUND
PROGRAMMABLE ANALOG
SIGNAL (-)
Signal description
START/STOP: input signal to start or stop the pump.
With the supplied cover connector the START/STOP (+)
signal is connected to the +24 Vdc pin and the
START/STOP (-) signal to the GROUND pin: in this
condition the pump automatically starts as soon as the
controller recognises the input supply ("Plug & Pump").
INTERLOCK: input signal to control the pump rotation.
With the supplied cover connector the INTERLOCK (+)
signal is connected to the +24 Vdc pin and the INTER-
LOCK (-)signal to the GROUND pin.
37
TECHNICAL INFORMATION
SIGNAL NAME
OUTPUT
87-901-001-01 (A)
INPUT/
IN
IN
IN
IN
IN
IN
IN
IN
OUT
OUT
OUT
OUT
OUT

Werbung