Error message
Fatal error
15 V DC supply missing
24 V DC supply missing
No Serial communication
Program version mismatch
Error
Air pressure too low
Disc Motor Overloaded
Dressing arm motor is stuck
Emergency Stop is active
Process has been stopped by lifting
protective shield
Pump Motor Overloaded
Removal rate too low, maximum
permitted time has been exceeded
Sample holder unable to move
upwards after process
Sample Motor Overloaded
AbraPlan-10
Instruction Manual
Explanation
No power to circuitry in AbraPlan-10
No power to circuitry in AbraPlan-10
Communication problem between
elements inside AbraPlan-10.
Incompatible software in
AbraPlan-10.
There may be a leak in the hose or
the compressor maybe defective.
Load pressure is too high.
This may be due to an internal
electrical problem.
Possible physical obstruction
blocking arm from moving.
The Emergency Stop is pushed in.
This is displayed until action is
taken.
The protective shield has been
lifted.
Pollutants in the coolant are
clogging the pump.
The machine has been grinding for
15 minutes without reaching the
goal.
There may be a problem with the
compressed air system.
There may be an internal electrical
problem.
Loading on the motor is too high.
Action required
Contact Technical Support.
Contact Technical Support.
Contact Technical Support.
Contact Technical Support.
Check the compressed air system
for possible cause.
Reduce force value.
Fatal – contact Technical Support.
Remove the obstruction.
Release the Emergency Stop.
Put the shield down into its
operating position.
Clean the cooling system and fill
with fresh coolant mixture.
When using Stock Removal, check
that the force was not altered during
operation.
Check the compressed air system.
This is fatal – contact Technical
Support.
Reduce the force value.