Disposal
CAUTION
The spent disposable CO
absorber cannot be prepared for
2
reuse.
At the end of life, dispose of the product in accordance with
applicable regulations.
It is recommended that the possibility of disposing of soda lime
in household waste be checked with local waste disposal
authorities.
In Europe, the applicable waste material code for soda lime is
EWC 160303/180106.
Technical data
Ambient conditions
During operation
Temperature
10 °C to 40 °C (50 °F to 104 °F)
Atmospheric
500 to 1200 hPa (7.3 psi to 17.4 psi)
pressure
Relative humidity
0 to 95 %
During storage and transport
Temperature
–20 °C to +50 °C (-4 °F to 122 °F)
CLIC adapter
–20 °C bis +55 °C (–4 °F bis 131 °F)
and intermediate
plate
Atmospheric
500 to 1200 hPa (7.3 psi to 17.4 psi)
pressure
Relative humidity
0 to 95 %
Resistance
Reduced by approx. 1 mbar in the bypass mode.
Depends on the breathing system of the
respective Dräger anesthesia workstation.
Compliance
Normal mode: approx. 1.3 mL/hPa
(1.3 mL/mbar)
Bypass function: approx. 0.4 mL/hPa
(0.4 mL/mbar)
Leakage
<5 mL/min
Classification
Class IIa
according to Directive 93/42/EEC, Appendix IX
UMDNS code
17-509
Universal Medical Device Nomenclature System
(Nomenclature for medical devices)
The medical device meet the requirements of the ISO 80601-2-13
standard.
List of accessories
CLIC Absorber Free
Infinity ID CLIC Absorber Free
Intermediate plate
Cleaning disc
CLIC adapter
Sealing
Sealing set for intermediate plate M 35 380
Instructions for use CLIC Absorber Free / Infinity ID CLIC Absorber Free / CLIC Adapter
MX 50 100
MX 50 120
MX 00 013
MX 50 081
MX 50 090
MX 50 072
Safety data sheet
The Drägersorb Free safety data sheet is available on request.