6. Disposal
Technical equipment and batteries do not belong in house-
hold waste. They must be disposed of at appropriate collec-
tion and disposal points.
7. Electromagnetic compatibility (EMC)
The unit is an electrical medical product and is subject to special pre-
cautionary measures with regard to EMC which must be published in
the instructions for use.
Portable and mobile HF communications equipment can affect the unit.
Use of the unit in conjunction with non-approved accessories can af-
fect the unit negatively and alter the electromagnetic compatibility.
The unit should not be used directly adjacent to or between other elec-
trical equipment.
The unit satisfies the EMC requirements of the international standard
IEC60601-1-2. The requirements are satisfied under the conditions de-
scribed in the tables below.
EN-50
Technical information
Technical information
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity test
IEC 60601 test
Compliance
level
level
Electrostatic
±6 kV contact
±6 kV contact
discharge
±8 kV air
±8 kV air
(ESD) IEC
61000-4-2
Electrical fast
±2 kV for
±2 kV for
transient/burst
power supply
power supply
IEC 61000-4-4
lines
lines
±1 kV for input/
±1 kV for input/
output lines
output lines
Surge IEC
±1 kV differen-
±1 kV differen-
61000-4-5
tial mode
tial mode
±2 kV common
±2 kV common
mode
mode
Voltage
<5 % U
(>95
<5 % U
T
dips, short
% dip in UT) for
% dip in UT) for
interruptions
0.5 cycle
0.5 cycle
and voltage
40 % U
(60 %
40 % U
T
variations on
dip in UT) for 5
dip in UT) for 5
power supply
cycles
cycles
input lines IEC
70 % U
(30 %
70 % U
T
61000-4-11
dip in UT) for
dip in UT) for
25 cycles
25 cycles
<5 % U
(>95
<5 % U
T
% dip in UT) for
% dip in UT) for
5 sec.
5 sec.
Power frequen-
3 A/m
3 A/m
cy (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
E
Electromagnetic environ-
ment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
(>95
Mains power quality
T
should be that of a typical
commercial or hospital
(60 %
environment. If the user
T
of the device requires
continued operation during
(30 %
power mains interruptions,
T
it is recommended that the
device is powered from
(>95
an uninterruptible power
T
supply or a battery.
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
EN-51