Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning with
immersion disinfection
Manual pre-cleaning
with brush and subse-
quent mechanical alka-
line cleaning and ther-
mal disinfection
Specific requirements
Cleaning brush: 30 mm/
∅: 4.5 mm, e.g. TA no. 011944
20 ml disposable syringe
When cleaning products with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
Drying phase: Use a lint-free cloth
or medical compressed air
Cleaning brush: 30 mm/
∅: 4.5 mm, e.g. TA no. 011944
20 ml disposable syringe
Place the instrument in a tray
that is suitable for cleaning
(avoiding rinsing blind spots).
Connect components with lumens
and channels directly to the rins-
ing port of the injector carriage.
Keep working ends open for
cleaning.
Place instruments in the tray with
their hinges open.
Reference
Chapter Manual cleaning/disinfection and
sub-chapter:
Chapter Manual cleaning with immer-
sion disinfection
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and sub-chap-
ter:
Chapter Manual pre-cleaning with a
brush
Chapter Mechanical alkaline cleaning
and thermal disinfecting