Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic DBS 37086 Implantationsanleitung Seite 9

Verlängerungsset für die tiefe hirnstimulation
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 37
Table 1. Device specifications for the Model 37086 Extension
Description
b,c
Resistance
Length
Distal end (lead)
Connector
Contact spacing
Lead entrance diameter
Outer diameter
Proximal end (neurostimulator)
Connector
Contact spacing
Diameter
a
All measurements are approximate.
b
Electrical resistance of this device only.
Resistance is proportional to length: long lengths have higher resistance that may limit
c
the amplitude.
Table 2. Material of components in the Model 37086 package
Component
Extension
Conductor wire
Distal end (lead)
Insulation
Setscrew connector block
Proximal end (neurostimulator)
Conductor wire insulation
Contacts
Insulation
Connector boots
Clear
White radiopaque
Neurostimulator plug
Setscrews
Model 37086
Maximum 38.0 Ω for all lengths
40 cm, 60 cm, or 95 cm
Quadripolar, in-line
4.3 mm
1.5 mm
3.8 mm
Octapolar, in-line
2.8 mm
1.3 mm
Material
Silver core MP35N
Silicone rubber
Stainless steel
Fluoropolymer
MP35N
Polyurethane
Silicone rubber
Silicone rubber, barium
sulfate
Polyurethane
Titanium
37086 2020-01-01  English 9
a
Material
contacts human
tissue
No
Yes
Yes
No
Yes
Yes
Yes
Yes
Yes
Yes

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis