8.

Maintenance

The display "Battery replacement recommended" on the Acculan 3Ti char-
ger GA677 (see TA022160) indicates when the battery should be replaced
due to reduced capacity.
Note
The "Battery replacement recommended" display on the Acculan 4 charger
GA320 has no function in the NiMH battery GA666 and GA676.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
9.

Troubleshooting

Replace product: have defective products repaired by Aesculap Techni-
cal Service, see Technical Service.
Issue
Detection
Battery not
No signal tone
working
when inserting the
battery into the
machine
Battery
Symbol "Battery
quickly empty
replacement rec-
ommended" illumi-
nates at the end of
the charging pro-
cess on the
Acculan 3Ti charger
GA677 (not on
Acculan 4 charger
GA320)

10. Storage and transport

Risk of injury from burns and damage to product in
case of short-circuit!
When storing the product, ensure that the prod-
WARNING
uct cannot short-circuit with other products or
metal objects.
In order to prevent damage during transport, ship product in the orig-
inal packaging and in an outer carton.
If the original packaging is no longer available, ensure the following
when packaging the product:
– The battery contacts are protected against short-circuit.
– The packaging completely encloses the product.
– The product is secured against movement within the packaging.
Completely charge the product upon receipt and before extended peri-
ods of storage.

11. Accessories/Replacement Parts

Art. no.
Designation
TA015514
Instructions for use, battery GA666, GA676 (A4)
TA011867
Instructions for use, battery GA666, GA676 (flyer)
6
Cause
Remedy
Battery defec-
Replace prod-
tive
uct.
Battery cells
Have battery
at end of life
cells replaced
by Aesculap
Technical Ser-
vice.

12. Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
Modifications carried out on medical technical equipment may result in
loss of warranty rights and applicable licenses.
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen /Germany
Phone:
+49 7461 95-1601
Fax:
+49 7461 16-939
E-Mail:
ats@aesculap.de
Or in the US:
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042
Aesculap Repair Hotline
Phone:
+1 800 214-3392
Fax:
+1 314 895-4420
Other service addresses can be obtained from the address indicated above.

13. Technical Data

13.1 Classification according to directive 93/42/EWG

Art. no.
Designation
GA666
Acculan battery
GA676
Acculan battery

13.2 Performance data, information and standards

Model
GA666
Cell type
NiMH
DC voltage
9.6 V
Capacity
1.05 Ah
Weight
approx. 0.304 kg
Dimensions (L x W x H)
approx.
120 mm x 43 mm x 5
0 mm
EMC
IEC/DIN 60601-1-2
Conforming to standard
IEC/DIN 60601-1
Class
I
I
GA676
NiMH
9.6 V
1.95 Ah
approx. 0.425 kg
approx. 140 mm x
43 mm x 50 mm
IEC/DIN 60601-1-2
IEC/DIN 60601-1