7.3
Reusable products
The service life of the product is limited by damage, normal wear, type and
duration of the application, handling, storage and transportation of the
product.
Careful visual and functional testing prior to next use is the best way to
identify a malfunctioning product.
7.4
Preparing for cleaning
Remove the battery from the Acculan machine.
7.5
Cleaning/Disinfection
Device-specific safety information on the reprocessing method
DANGER
Risk of electric shock and fire hazard!
Do not use flammable or explosive cleaning or disinfecting solu-
tions.
Ensure that no fluids will penetrate the product.
7.6
Wipe disinfection of electrical equipment
Phase
Step
I
Cleaning
II
Wipe disinfection
RT:
Room temperature
*
Recommended: Meliseptol® Wipes sensitive (B. Braun)
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Phase II
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe contact time (at least 3 min).
7.7

Inspection, maintenance and checks

Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
Set aside the product if it is damaged.
8.
Maintenance
The display "Battery replacement recommended" on the Acculan 4 charger
GA320 (see TA014534/TA014535) indicates when the battery should be
replaced due to reduced capacity.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
T
t
Conc.
[°C/°F]
[min]
[%]
RT
0.5
-
≥3
RT
-
CAUTION
Damage to, or destruction of the product caused by mechanical clean-
ing/disinfection!
Only clean and disinfect the product manually.
Do not sterilize the product under any circumstances.
CAUTION
Risk of damage to the product due to inappropriate cleaning/disin-
fecting agents!
Only use cleaning/disinfecting agents approved for surface clean-
ing. Follow the manufacturer's instructions for the respective
cleaning/disinfecting agent.
Do not introduce any fluid onto the electrical contacts of the bat-
tery.
Water quality
Chemistry
-
17 % 1-Propanol,
0.23 % Didecyldimethylammonium chloride*
-
17 % 1-Propanol,
0.23 % Didecyldimethylammonium chloride*
en
5