General information

2.1

Use/intended use:

Task/Function
The Li-Ion batteries serve to supply energy to the Acculan machines and
include the activation and control of the electric motors of the Acculan
machines according to user commands.
Application Environment
The product fulfills the requirements of Type BF in accordance with
IEC/DIN EN 60601-1.
For use in operating rooms, outside of the explosive hazard zone (e.g. areas
with highly pure oxygen or anesthetic gases).
2.2

Main functions and design characteristics

Art. no.
DC voltage
GA666
9.6 V
GA676
9.6 V
2.3

Indications

The indications can be found in the instructions for use for the correspond-
ing Acculan machines.
2.4

Contraindications

The contraindications can be found in the instructions for use for the cor-
responding Acculan machines.
3.

Safe handling

CAUTION
Federal law restricts this device to sale by, or on order of a physician!
Risk of injury and damage to property due to
improper handling of the product!
Comply with instructions for use for the
WARNING
Acculan 4 machines and the Acculan 3Ti
machines.
Follow the instructions for use of all products
used.
Fire and explosion hazard due to battery poles
shorted by fluids or metal components!
Do not short-circuit, open, crush or incinerate
WARNING
battery.
Damage to or destruction of the batteries due to
reprocessing!
Do not sterilize batteries.
CAUTION
Inspect the new product after removing its transport packaging and
prior to first use to ensure it is in good working order; charge with char-
ger GA320 or GA677 and verify functionality.
To prevent damage caused by improper setup or operation, and to not
compromise the manufacturer warranty and liability:
– Use the product according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
– Follow the application advisories acc. to standard, see Extracts from
relevant standards.
Capacity
1.05 Ah
1.95 Ah
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
Keep the instructions for use accessible for the user.
Always adhere to applicable standards.
Do not use the product if it is damaged or defective.
Set aside the product if it is damaged.
Store product at room temperature (Maximum recommended duration
of storage is 3 months; keep out of direct sunlight).
Keep product dry.
Observe the instructions for use for the corresponding Acculan
machines.
Use battery regularly and always store in the charger.
4.

Product description

4.1

Scope of Supply

Art. no.
Designation
GA666
Acculan NiMH battery short
– or –
GA676
Acculan NiMH battery
TA011867
Instructions for use, Acculan battery GA666, GA676
4.2

Components required to operate the device

Acculan 4 charger GA320 or Acculan 3Ti charger GA677
Acculan 4 or Acculan 3Ti machine:
Art. no.
Designation
GA330
Acculan 4 Drill and reamer
GA331
Acculan 4 Oscillating saw
GA334
Acculan 4 Reciprocating saw
GA336
Acculan 4 Oscillating saw straight
GA337
Acculan 4 Oscillating sternum saw
GA340
Acculan 4 Dermatome
GA341
Acculan 4 Dermatome 0.1 mm
GA344
Acculan 4 Small drill
GA668
Acculan 3Ti Oscillating saw straight
GA670
Acculan 3Ti Dermatome
GA671
Acculan 3Ti Small drill
GA672
Acculan 3Ti Drill and reamer
GA673
Acculan 3Ti Oscillating saw
GA674
Acculan 3Ti Reciprocating saw
GT666
Acculan 4 TPLO saw
GT668
Acculan 3Ti TPLO saw